E.coli O157: H7 PCR detectio ornamentum
Product Name
E.Coli O157: H7 PCR Deprehensio Kit (Lyophilized)
Magnitudo
48tests/ornamentum, 50tests/kit
Intentum usum
Escherichia coli O157:H7 (E.coli O157:H7) est gram-negativum bacterium ad genus Enterobacteriaceae pertinens, quod magnam copiam Vero toxin producit.Clinocally solet repente cum gravi abdominis dolor fluxus fluxus post hemorrhagic fluxus paucos dies, quae potest ducere febricitantem et non febricitantem et mors gravi casibus.Hoc ornamentum est idoneam ad qualita deprehendatur Escherichia coli O157: H7 in cibo, aqua exempla, faeces, vomit, Bacteria, enhancing Liquid et alias exempla per principium est realis-vicis PCR.The ( Lyophilized), quod habet DNA amplificare enzyme, reactionem quiddam, specifica primers et probes requiritur ad fluorescent PCR deprehendatur ..
Product Contents
Components | sarcina | specificatio | Ingredient |
E.Coli O157: H7 PCR Nobis | I utrem (pulveris Lyophilized) | 50Test | dNTPs, MgCl2, Primarios, Probes, Reverse Transcriptase, Taq DNA polymerase |
6×0.2ml 8 tubus bene habena(Lyophilized) | 48Test | ||
Positivum Imperium | 1*0.2ml tubus (lyophilized) | 10Tests | Plasmid quibus E.Coli O157: H7 certis fragmentorum |
dissolutio solutionis | 1.5 ml Cryotube | 500uL | / |
Imperium negativum | 1.5 ml Cryotube | 200uL | 0.9% NaCl |
Repono & Shelf Life
(1) Ornamentum ad locus temperatus transportari potest.
(2) Pluteae vita est duodeviginti mensium -20℃ et 12 menses ante 2℃~30℃.
(3) Vide pittacium in ornamento ad diem producendi et exspirandi.
(4) Pulvis lyophilisatus iodi versionis in -20℃ post dissolutionem condi debet et repetita duratus -thaw minus quam 4 temporibus esse debet.
Instrumenta
GENECHECKER UF-150, UF-300 real-time fluorescens PCR instrumentum.
Operatio Diagram
a) Utrem versionem:
b) 8 tubus litera bene habena;
Per amplificationem
CommendaturOccasum
Gradus | Cycle | Temperatus (℃) | Tempus | Fluorescens alveum |
1 | 1 | 95 | 2min | |
2 | 40 | 95 | 5s | |
60 | 10s | Collect FAM fluorescens |
* Nota: et signo Fluorescence channel et colligi ad LX ℃.
INTERPRETATIO Test Results
Channel | Proventuum interpretatio |
FAM Channel | |
Ct≤35 | E.Coli O157: H7 positive |
Undet | E.Coli O157: H7 negans |
35 | Suspicious resut, retest* |
*Si retest effectus FAM canalis habet Ct valorem ≤40 et typicam "S" figuram amplificationis curvae ostendit, effectus positivus interpretatur, alioquin negativus est.